Enantiomeric Separation of S-Epichlorohydrin and R-Epichlorohydrin by Capillary Gas Chromatography with FID Detector
- 1 Dr. Reddy’s Laboratories Ltd. Active Pharmaceutical Ingredients, IPDO, Hyderabad, India
- 2 Department of Pediatrics, SUNY Stony Brook Medical Center, Stony Brook, New York, USA
- 3 Dr. Reddy’s Laboratories Ltd. Active Pharmaceutical Ingredients, IPDO, Hyderabad, India
- 4 Department of Chemistry, J.N.T. University, Hyderabad, India
- 5 Dr. Reddy’s Laboratories Ltd. Active Pharmaceutical Ingredients, IPDO, Hyderabad, India
Abstract
The aim of this study was to develop a simple and derivatization free method for the Quantification of S-Epichlorohydrin in R-Epichlorohydrin by using a gas chromatography coupled with flame ionization detector (FID). Enantiopure epichlorohydrin was a valuable epoxide key starting material for preparing optically active Rivaroxaban. The enantiomeric separations of S-Epichlorohydrin and R-Epichlorohydrin were achieved on Gamaa-Dex-225 (30 meters × 0.25 mm I.D, 0.25 μm) column with a total run time of 30 min. Nitrogen was used as a carrier gas with constant pressure 25.0 psi. The critical experimental parameters such as, column selection, flow rate, injection volume and diluent were studied and optimized. Excellent correlation coeffient between peak responses and concentrations was >0.9998. The recoveries of S-Epichlorohydrin spiked in R-Epichlorohydrin were in the range from 98.2% to 102.8%. Limit of quantitation for S-Epichlorohydrin was sufficiently lower than limits specified by ICH. The method has validated as per International Conference on Harmonization (ICH) guidelines. A precise, accurate, linear and robust Gas Chromatography method was developed for the quantification of S-Epichlorohydrin in R-Epichlorohydrin for Rivaroxaban.
- Lehmann, R., Voelter, W. and Liebich, H.M. (1997) Capillary Electrophoresis in Clinical Chemistry. Journal of Chromatography B, 697, 3-35. http://dx.doi.org/10.1016/S0378-4347(97)00183-7
- Naylor, S., Benson, L.M. and Tomlinson, A.J. (1996) Application of Capillary Electrophoresis and Related Techniques to Drug Metabolism Studies. Journal of Chromatography A, 735, 415-438. http://dx.doi.org/10.1016/0021-9673(96)00068-4
- United States Environmental Protection Agency, Office of Air Quality Planning and Standards (1984) Final Report, EPA-450/4-84-007j.
- World Health Organization (1992) Chemical Review: Epichlorohydrin. Dangerous Properties of Industrial Materials Report, 12, 150-170.
- International Agency for Research on Cancer (1999) Re-Evaluation of Some Organic Chemicals, Hydrazine and Hydrogen Peroxide. Monographs on the Valuation of Carcinogenic Risk to Humans, 71, 603.
- Susanne, R., Alexander, S. and Jens, P. (2005) Discovery of Novel Antithrombotic Agent BAY 59-7939. Journal of Medicinal Chemistry, 48, 5900-5908.
- Perzborn, E., Strassburger, J. and Wilmen, A. (2005) In Vitro and In Vivo Studies of BAY 59-7939. Journal of Thrombosis and Haemostasis, 3, 514-521. http://dx.doi.org/10.1111/j.1538-7836.2005.01166.x
- Lasa, M., Garcia, R. and Millán, E. (2006) A Convenient Method for Epichlorohydrin Determination in Water Using Headspace-Solid-Phase Micro Extraction and Gas Chromatography. Journal of Chromatographic Science, 44, 438-443. http://dx.doi.org/10.1093/chromsci/44.7.438
- Shaik, J.V., Ganduri, R.B. and Sait, S. (2011) Estimation of Epichlorohydrin Content in Pharmaceutical Drug Substances by Capillary Gas Chromatography with Flame Ionisation Detection. Journal of Chemical and Pharmaceutical Research, 3, 392-399.
- Kadiyala, R.V., Kothapalli, P.K., Peddolla, M.R., Rajput, P., Sharma, H.K., Budeti, S.R., Gandham, H. and Nowduri, A. (2014) Development and Validation of a Gas Chromatography Method for the Trace Level Determination of Allylamine in Sevelamer Hydrochloride and Sevelamer Carbonate Drug Substances. Scientia Pharmaceutica, 82, 117-128. http://dx.doi.org/10.3797/scipharm.1309-12
- Karthikeyan, K., Arularasu, G.T., Perumalsamy, D. and Chandrasekara, P.K. (2010) Determination of Residual Epichlorohydrin in Sevelamer Hydrochloride by Static Headspace Gas Chromatography with Flame Ionization Detection. Scientia Pharmaceutica, 78, 835-846. http://dx.doi.org/10.3797/scipharm.1007-20