An Ultra-High Performance Liquid Chromatography-Tandem Mass Spectrometry Method for the Quantification of Vancomycin Requiring Only 2 µL of Rabbit Serum
- 1 Faculty of Pharmaceutical Sciences, The University of British Columbia, Vancouver, Canada
- 2 Faculty of Pharmaceutical Sciences, The University of British Columbia, Vancouver, Canada
- 3 Faculty of Pharmaceutical Sciences, The University of British Columbia, Vancouver, Canada
- 4 Faculty of Pharmaceutical Sciences, The University of British Columbia, Vancouver, Canada
Abstract
A highly sensitive ultra-high performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS) method was developed for the quantification of vancomycin (VAN) in low volumes of rabbit serum. For each analysis, 2 μL rabbit serum was precipitated with methanol that contained the internal standard teicoplanin (TEI). The supernatant was transferred into a 384 well-plate, diluted with water, covered with a pierceable silicone mat and 5 μL was analyzed in positive ionization mode. The UHPLC-MS/MS consisted of an Agilent 1290 Infinity UHPLC system connected to an AB Sciex QTrap ® 5500 hybrid linear ion-trap triple quadrupole mass spectrometer equipped with a Turbo Spray source. Chromatographic separation was achieved using a Waters Acquity UPLC BEH C 18 (1.7 μm, 2.1 mm × 100 mm) column, a VanGuard (1.7 μm, 2.1 × 5 mm) guard column and a mobile phase of water and methanol both containing 5 mM ammonium acetate with 0.1% formic acid. VAN was quantified with multiple reaction monitoring using the transitions of m/z 725.5/144.2, and TEI was monitored at m/z 940.6/316.4. The accuracy, precision, linearity, range and lower limit of quantification (LLOQ) were determined. The accuracy was ≤9.93% and the precision was ≤10.6%. The range was established as 0.1 to 40 μg·mL -1 . The LLOQ was 0.1 μg·mL -1 VAN requiring 2 μL of sample with an accuracy of -20.2% and precision of 8.39%. The method was applied successfully to determine the VAN concentrations in rabbit serum after the i.v. administration of VAN via implanted ear catheters.
- Cao, Y., Yu, J., Chen, Y., Zhang, J., Wu, X., Zhang, Y. and Li, G. (2014) Development and Validation of a New Ultra-Performance Liquid Chromatographic Method for Vancomycin Assay in Serum and Its Application to Therapeutic Drug Monitoring. Therapeutic Drug Monitoring, 36, 175-181. https://doi.org/10.1097/FTD.0b013e3182a458bc
- Li, X., Wang, F., Xu, B., Yu, X., Yang, Y., Zhang, L. and Li, H. (2014) Determination of the Free and Total Concentrations of Vancomycin by Two-Dimensional Liquid Chromatography and Its Application in Elderly Patients. Journal of Chromatography B Analytical Technologies in the Biomedical and Life Sciences, 969, 181-189. https://doi.org/10.1016/j.jchromb.2014.08.002
- Pfaller, M.A., Krogstad, D.J., Granich, G.G. and Murray, P.R. (1984) Laboratory Evaluation of Five Assay Methods for Vancomycin: Bioassay, High-Pressure Liquid Chromatography, Fluorescence Polarization Immunoassay, Radioimmunoassay, and Fluorescence Immunoassay. Journal of Clinical Microbiology, 20, 311-316.
- Gunther, M., Saxinger, L., Gray, M. and Legatt, D. (2013) Two Suspected Cases of Immunoglobulin-Mediated Interference Causing Falsely Low Vancomycin Concentrations with the Beckman PETINIA Method. The Annals in Pharmacotherapy, 47, e19. https://doi.org/10.1345/aph.1R566
- Hagihara, M., Sutherland, C. and Nicolau, D.P. (2013) Development of HPLC Methods for the Determination of Vancomycin in Human Plasma, Mouse Serum and Bronchoalveolar Lavage Fluid. Journal of Chromatographic Science, 51, 201-207. https://doi.org/10.1093/chromsci/bms128
- Konig, K., Kobold, U., Fink, G., Leinenbach, A., Dulffer, T., Thiele, R., Zander, J. and Vogeser, M. (2013) Quantification of Vancomycin in Human Serum by LC- MS/MS. Clinical Chemistry and Laboratory Medicine, 51, 1761-1769. https://doi.org/10.1515/cclm-2013-0142
- Cheng, C., Liu, S., Xiao, D., Hollembaek, J., Yao, L., Lin, J. and Hansel, S. (2010) LC-MS/MS Method Development and Validation for the Determination of Polymyxins and Vancomycin in Rat Plasma. Journal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences, 878, 2831-2838. https://doi.org/10.1016/j.jchromb.2010.08.037
- Tsai, I.L., Sun, H.Y., Chen, G.Y., Lin, S.W. and Kuo, C.H. (2013) Simultaneous Quantification of Antimicrobial Agents for Multidrug-Resistant Bacterial Infections in Human Plasma by Ultra-High-Pressure Liquid Chromatography-Tandem Mass Spectrometry. Talanta, 116, 593-603. https://doi.org/10.1016/j.talanta.2013.07.043
- Cazorla-Reyes, R., Romero-Gonzalez, R., Frenich, A.G., Rodriguez Maresca, M.A. and Martinez Vidal, J.L. (2014) Simultaneous Analysis of Antibiotics in Biological Samples by Ultra High Performance Liquid Chromatography-Tandem Mass Spectrometry. Journal of Pharmaceutical and Biomedical Analysis, 89, 203-212. https://doi.org/10.1016/j.jpba.2013.11.004