Perioperative Nursing Management of an Animal Model of Axial Flow Blood Pump Implantation
- 1 Wuhan Asian Heart Disease Hospital Affiliated to Wuhan University of Science and Technology, Wuhan, China
- 2 Medical College of Wuhan University of Science and Technology, Wuhan, China
- 3 Wuhan Asian Heart Disease Hospital Affiliated to Wuhan University of Science and Technology, Wuhan, China
- 4 Wuhan Asian Heart Disease Hospital Affiliated to Wuhan University of Science and Technology, Wuhan, China
- 5 Huazhong Agricultural University, Wuhan, China
- 6 Wuhan Asian Heart Disease Hospital Affiliated to Wuhan University of Science and Technology, Wuhan, China
Abstract
This study aimed to explore the characteristics of perioperative nursing of experimental goats using self-made axial-flow blood pump implantation and provided theoretical nursing knowledge and practice-based evidence for the clinical application of domestically manufactured artificial cardiac pumps. Methods: Seven experimental goats were used in this study, three for pre-testing and four for the formal experiments. According to the surgical requirements for axial-flow blood pump implantation into the cardiac apex, we creatively designed and made a series of highly practical animal surgical instruments including a composite disassemblable bed for experimental animal transferring and monitoring, a multifunctional animal surgery bed, and portable medical supporting equipment. We also applied for two national invention patents and one utility model patent. Active measures were taken to ensure careful preparation before surgery, close collaboration during surgery, and effective management of complications after surgery. Results: Two of the four experimental goats died during surgery because of a massive hemorrhage caused by distal anastomotic failure and air embolism-induced cardiac arrest caused by air leakage from the outlet into the heart due to poor connection of the auxiliary pressure tap (used to measure left ventricular pressure). The mean survival time of the remaining three experimental goats was 22.7 hours. Conclusion: This study was the first to systematically and comprehensively investigate the perioperative nursing management of axial-flow blood pump implantation using animal models. These findings could greatly promote further clinical applied nursing research of self-made artificial cardiac pump implantation in experimental goats.
- El Banayosy, A., Fey, O., Sarnowski, P., et al. (2001) Midterm Follow up of Patients Discharged from Hospital under Left Ventricular Assistance. Journal of Heart and Lung Transplantation, 20, 53-58.
- McCarthy, P.M. (1995) Heart Mate Implantable Left Ventricular Assist Device: Bridge to Transplantation and Future Applications. The Annals of Thoracic Surgery, 59, 46-51.
- Bank, A.J., Mir, S.H., Nguyen, D.Q., et al. (2000) Effects of Left Ventricular Assist Devices on Outcomes in Patients Undergoing Heart Transplantation. The Annals of Thoracic Surgery, 69, 1369-1374.
- Hsu, R.B., Chu, S.H. and Chien, C.Y. (2000) Heart Mate Left Ventricular Assist Device for Long-Term Circulatory Support. Journal of the Formosan Medical Association, 99, 336-340.
- Koul, B., Solem, J.O., Steen, S., et al. (1998) Heart Mate Left Ventricu1ar Assist Device as Bridge to Heart Transplantation. The Annals of Thoracic Surgery, 65, 1625-1630.
- Rogers, J.G., Butler, J., Lansman, S.L., et al. (2007) Chronic Mechanical Circulatory Support for Inotrope-Dependent Heart Failure Patients Who Are Not Trans-Plant Candidates: Results of the Intrepid Trial. Journal of the American College of Cardiology, 50, 741-747. http://dx.doi.org/10.1016/j.jacc.2007.03.063
- Patel, S.M., Throckmorton, A.L., Untaroiu, A., et al. (2005) The Status of Failure and Reliability Testing of Artificial Blood Pumps. ASAIO Journal, 51, 440-451. http://dx.doi.org/10.1097/01.mat.0000169083.90253.3c
- Farrar, D.J., Bourque, K., Dague, C.P., et al. (2007) Design Features, Developmental Status, and Experimental Results with the Heartmate Ⅲ Centrifugal Left Ventricular Assist System with a Magnetically Levitated Rotor. ASAIO Journal, 53, 310-315. http://dx.doi.org/10.1097/MAT.0b013e3180536694
- Wieselthaler, G.M.O., Driscoll, G., Jansz, P., et al. (2010) Initial Clinical Experience with a Novel Left Ventricular Assist Device with a Magnetically Levitated Rotor in a Multiinstitutional Trial. Journal of Heart and Lung Transplantation, 29, 1218-1225. http://dx.doi.org/10.1016/j.healun.2010.05.016
- Morshuis, M., Schoenbrodt, M., Nojiri, C., et al. (2010) Dura Heart Magnetically Levitated Centrifugal Left Ventricular Assist System for Advanced Heart Failure Patients. Expert Review of Medical Devices, 7, 173-183. http://dx.doi.org/10.1586/erd.09.68
- Zhang, Y., Hu, S.S., Zhou, J.Y., et al. (2009) The Study of FW Type of Axial Flow Pump in Vitro Hemolysis and Animal Experiment. Chinese Journal of Clinical Thoracic and Cardiovascular Surgery, 16, 114-117. (In China)