Evaluation of Primary Stability of Reverse Shoulder Arthroplasty Based on Finite Element Analysis
- 1 Department of Medical Engineering and Technology, School of Allied Health Sciences, Kitasato University, Sagamihara, Japan
- 2 Graduate School of Medical Sciences, Kitasato University, Sagamihara, Japan
- 3 Graduate School of Medical Sciences, Kitasato University, Sagamihara, Japan
- 4 Graduate School of Medical Sciences, Kitasato University, Sagamihara, Japan
- 5 Department of Medical Engineering and Technology, School of Allied Health Sciences, Kitasato University, Sagamihara, Japan
- 6 Department of Medical Engineering and Technology, School of Allied Health Sciences, Kitasato University, Sagamihara, Japan
- 7 Department of Orthopedic Surgery, School of Medicine, Kitasato University, Sagamihara, Japan
Abstract
As the number of reverse shoulder arthroplasty (RSA) procedures increases, so does the incidence of revision surgery. Baseplate-related complications account for the highest proportion of these revisions, and it has been reported that improving baseplate fixation reduces the likelihood of failure. The present study aims to evaluate the initial stability of the baseplate to the glenoid after RSA. A finite element analysis (FEA) was performed using LS-DYNA models of the scapula and the SMR shoulder system, with a load of 30 N applied in both abduction and flexion, using the baseplate implantation surface as the reference. Micromotion was defined as the difference in displacement between the baseplate and the scapular fossa. The results demonstrated that micromotion between the glenoid and the baseplate diminished with increasing elevation in both abduction and flexion. It is hypothesised that in the SMR shoulder system, the screws are pressed into the glenoid during abduction, thereby contributing to enhanced initial stability.
- Parada, S.A., Flurin, P., Wright, T.W., Zuckerman, J.D., Elwell, J.A., Roche, C.P., et al . (2021) Comparison of Complication Types and Rates Associated with Anatomic and Reverse Total Shoulder Arthroplasty. Journal of Shoulder and Elbow Surgery , 30, 811-818. https://doi.org/10.1016/j.jse.2020.07.028
- Stephens, B.C., Simon, P., Clark, R.E., Christmas, K.N., Stone, G.P., Lorenzetti, A.J., et al . (2016) Revision for a Failed Reverse: A 12-Year Review of a Lateralized Implant. Journal of Shoulder and Elbow Surgery , 25, e115-e124. https://doi.org/10.1016/j.jse.2015.09.027
- Bonnevialle, N., Geais, L., Müller, J.H. and Berhouet, J. (2020) Effect of RSA Glenoid Baseplate Central Fixation on Micromotion and Bone Stress. JSES International , 4, 979-986. https://doi.org/10.1016/j.jseint.2020.07.004
- Hopkins, A.R., Hansen, U.N., Bull, A.M.J., Emery, R. and Amis, A.A. (2008) Fixation of the Reversed Shoulder Prosthesis. Journal of Shoulder and Elbow Surgery , 17, 974-980. https://doi.org/10.1016/j.jse.2008.04.012
- Zhang, M., Junaid, S., Gregory, T., Hansen, U. and Cheng, C. (2019) Effect of Baseplate Positioning on Fixation of Reverse Total Shoulder Arthroplasty. Clinical Biomechanics , 62, 15-22. https://doi.org/10.1016/j.clinbiomech.2018.12.021
- Boileau, P., Watkinson, D.J., Hatzidakis, A.M. and Balg, F. (2005) Grammont Reverse Prosthesis: Design, Rationale, and Biomechanics. Journal of Shoulder and Elbow Surgery , 14, S147-S161. https://doi.org/10.1016/j.jse.2004.10.006
- Ackland, D.C., Patel, M. and Knox, D. (2015) Prosthesis Design and Placement in Reverse Total Shoulder Arthroplasty. Journal of Orthopaedic Surgery and Research , 10, Article No. 101. https://doi.org/10.1186/s13018-015-0244-2
- Farron, A., Terrier, A. and Büchler, P. (2006) Risks of Loosening of a Prosthetic Glenoid Implanted in Retroversion. Journal of Shoulder and Elbow Surgery , 15, 521-526. https://doi.org/10.1016/j.jse.2005.10.003
- Friedman, R.J., Sun, S., She, X., Esposito, J., Eichinger, J.K. and Yao, H. (2021) Effects of Increased Retroversion Angle on Glenoid Baseplate Fixation in Reverse Total Shoulder Arthroplasty: A Finite Element Analysis. Seminars in Arthroplasty : JSES , 31, 209-216. https://doi.org/10.1053/j.sart.2020.11.014
- Sanjay, G. and Prosenjit, D. (2004) Bone Geometry and Mechanical Properties of the Human Scapula Using Computed Tomography Data. Trends in Biomaterials and Artificial Organs , 17, 61-69.