Abstract
Our aim was to evaluate the potential use of BioMedFlex® (BMF), a new resilient, hard-carbon, thin- film coating, as a blood journal bearing material in Cleveland Heart’s continuous-flow left and right ven- tricular assist devices (VADs). BMF is not classified as a diamond-like carbon (DLC) and differs from other thin-film carbon coatings by its high flexural strength, radiopacity, and wear resistance. A 2- to 4-μm-thick BMF adhesion layer was deposited on the VAD journal bearing surfaces. A commercial DLC coating used in other clinical blood pump applications was used as a control. Durability and reliability of the BMF coating was verified in severe pump start/stop testing using 20 BMF-coated journal bearing pairs. The BMF-coated surfaces showed no coating failures, whereas 57% of the DLC bearing pairs developed scratches through the carbon coating, documenting that BMF can provide a durable coating in our blood journal bearing application. In conclusion, BMF has shown qualities that support its significant advantages as an alternative journal bea- ring material in Cleveland Heart pumps. Our plan includes biocompatibility testing with ongoing animal studies, endurance testing with submerged pumps running in saline, and assessment of batch coating processing capability.
- Fukamachi, K., Horvath, D.J., Massiello, A.L., Ootaki, Y., Kamohara, K., Akiyama, M., Zahr, F., Kopcak, M.W., Dessoffy, R., Chen, J.F., Benefit, S. and Golding, L.R. (2005) Development of a small implantable right ventricular assist device. American Society for Artificial Internal Organs Journal, 51(6), 730-735.
- Ootaki, Y., Kamohara, K., Akiyama, M., Zahr, F., Kopcak, M.W., Dessoffy, R., Massiello, A.L., Horvath, D.J., Chen, J.F., Benefit, S., Golding, L.R. and Fukamachi, K. (2005) Initial in vivo evaluation of the DexAide right ventricular assist device. American Society for Artificial Internal Organs Journal, 51(6), 739-742.
- BioMedFlex, LLC, Accessed November 27, 2009. http:// www.BioMedFlex.com
- Yamazaki, K., Litwak, P., Tagusari, O., Mori, T., Kono, K., Kameneva, M., Watach, M., Gordon, L., Miyagishima, M., Tomioka, J., Umezu, M., Outa, E., Antaki, J.F., Kormos, R.L., Koyanagi, H. and Griffith, B.P. (1998) An implantable centrifugal blood pump with a recirculating purge system (Cool-Seal system). Artificial Organs, 22(6), 466-474.
- Malanoski, S.B., Belawski, H., Horvath, D.J., Smith, W.A. and Golding, L.R. (1998) Stable blood lubricated hydrodynamic journal bearing with magnetic loading. American Society for Artificial Internal Organs Journal, 44(5), M737-M740.
- Takaseya, T., Fumoto, H., Shiose, A., Arakawa, Y., Rao, S., Horvath, D.J., Massiello, A.L., Mielke, N.A., Chen, J.F., Zhou, Q., Dessoffy, R., Kramer, L., Benefit, S., Golding, L.R. and Fukamachi, K. (2009) In vivo biocompatibility evaluation of a new resilient, hard-carbon, thin-film coating for ventricular assist devices. Artificial Organs (in Press).