An investigation using high-precision CCD laser displacement sensor to measure body surface motion induced by heartbeat
- 1 Department of Mechanical Engineering, Faculty of Engineering Science, Kansai University, Osaka, Japan
- 2 Department of Management Systems Engineering, Faculty of System Design, Tokyo Metropolitan University, Tokyo, Japan
- 3 Department of Mechanical Engineering, Faculty of Engineering Science, Kansai University, Osaka, Japan
- 4 Department of Mechanical Engineering, Faculty of Engineering Science, Kansai University, Osaka, Japan
Abstract
The aim of this study was to measure the motion of the body surface induced by heartbeat by using a high-precision laser sensor in order to provide a reference for other research on medical remote sensing techniques that are unconstrained and non-contact. Body surface motions at some points on the chest and on the dorsal region were measured for eight men with a laser displacement sensor. For comparison, an electrocardiograph was also used to monitor the heartbeat. As a result, we confirmed cyclic motion of about 0.1 mm in amplitude everywhere on the body surface, not only from the front of the body but also from the back and the neck. In contrast, which part on the body was most suitable for measurement had to be taken into account because a difference in motion amplitudes was observed in different parts on the body. The amplitude of the motions also depended on the conditions of the body surface as well as underneath the skin. These results show the possibility of acquiring information about heartbeats from anywhere on the body by using an unconscious sensing technique, and the potential for the technique to monitor the condition of the body and personal physical characteristics.
- Ciaccio, E.J., Hiatt, M., Hegyi, T. and Drzewiecki, G.M. (2007) Measurement and monitoring of electrocardiogram belt tension in premature infants for assessment of respiratory function. Biomedical Engineering Online, 6, 1-11.
- Jacobs, J., Embree, P., Glei, M., Christensen, S. and Sullivan, P. (2004) Characterization of a novel heart and respiratory rate sensor. Conference Proceedings IEEE Engineering in Medicine and Biology Society, San Francisco, 1-5 September 2004, 2223-2226.
- Wang, F., Tanaka, M. and Chonan, S. (2006) Development of a wearable mental stress evaluation system using PVDF film sensor. Journal of Advanced Science, 18, 170-173. doi:10.2978/jsas.18.170
- Chen, K.-M., Huang, Y., Zhang, J.P. and Norman, A. (2000) Microwave life-detection systems for searching human subjects under earthquake rubble or behind barrier. IEEE Transactions on Biomedical Engineering, 47, 105-114. doi:10.1109/10.817625
- Chen, K.-M., Misra, D., Wang, H., Chuang, H.-R. and Postow, E. (1986) An X-band microwave life-detection system. IEEE Transactions on Biomedical Engineering, 33, 697-702. doi:10.1109/TBME.1986.325760
- Uenoyama, M., Matsui, T., Yamada, K., Suzuki, S., Takase, B., Suzuki, S. and Ishihara, M. (2006) Non-contact respiratory monitoring system using a ceiling-attached micro-wave antenna. Medical and Biological Engineering and Computing, 44, 835-840. doi:10.1007/s11517-006-0091-8
- Suzuki, S., Matsui, T., Kawahara, H. and Gotoh, S. (2009) Development of a noncontact and long-term respiration monitoring system using microwave radar for hibernating black bear. Zoo Biology, 28, 259-270. doi:10.1002/zoo.20229
- Suzuki, S., Matsui, T., Imuta, H., Uenoyama, M., Yura, H., Ishihara, M. and Kawakami M. (2008) A novel autonomic activation measurement method for stress monitoring: Non-contact measurement of heart rate variability using a compact microwave radar. Medical and Biological Engineering and Computing, 46, 709-714. doi:10.1007/s11517-007-0298-3
- Droitcour, A., Lubecke, V., Lin, J. and Boric-Lubecke, O. (2001) A microwave radio for Doppler radar sensing of vital signs. IEEE MTT-S International Proceedings of Microwave Symposium Digest, Phoenix, 20-25 May 2001, 175-178.
- Lin, J.C. (1992) Microwave sensing of physiological movement and volume change: A review. Bioelectromagnetics, 13, 557-565. doi:10.1002/bem.2250130610
- Li, C. and Lin, J. (2008) Random body movement cancellation in Doppler radar vital sign detection. IEEE Transactions on Microwave Theory and Techniques, 56, 3143-3152. doi:10.1109/TMTT.2008.2007139