Comparative Study of Imaging Characteristics of I-125 Imaging Using the Siemens Inveon Scanner and Siemens Symbia TruePoint
- 1 Molecular Imaging Research Center, Korea Cancer Center Hospital, Korea Institute of Radiological and Medical Sciences (KIRAMS), Seoul, Korea
- 2 Department of Nuclear Medicine, Korea Cancer Center Hospital, Korea Institute of Radiological and Medical Sciences (KIRAMS), Seoul, Korea
- 3 Molecular Imaging Research Center, Korea Cancer Center Hospital, Korea Institute of Radiological and Medical Sciences (KIRAMS), Seoul, Korea
- 4 Department of Nuclear Medicine, Korea Cancer Center Hospital, Korea Institute of Radiological and Medical Sciences (KIRAMS), Seoul, Korea
- 5 Molecular Imaging Research Center, Korea Cancer Center Hospital, Korea Institute of Radiological and Medical Sciences (KIRAMS), Seoul, Korea
- 6 Molecular Imaging Research Center, Korea Cancer Center Hospital, Korea Institute of Radiological and Medical Sciences (KIRAMS), Seoul, Korea
- 7 Korea Drug Development Platform Using Radio-Isotope (KDePRI), Seoul, Korea
- 8 Radiologcial and Medico-Oncological Sciences, University of Science and Technology (UST), Seoul, Korea
Abstract
Objective: Although Iodine-125 ( 125 I) has been widely used for in vitro studies because of its relatively long half-life (60.1 days), 125 I imaging is limited because of its low energy (27 - 35 keV), even in an animal-dedicated system. In this study, imaging characteristics of 125 I were assessed using a small animal-dedicated imaging system and clinical scanner. Methods: Using the Siemens Inveon and Siemens Symbia TruePoint systems, imaging characteristics such as resolution, sensitivity, and image quality were compared. Mouse high resolution (MHR-0.5), mouse general purpose (MGP-1.0), and mouse high sensitivity (MHS-2.0) collimators were used for the Inveon scanner, and low energy high-resolution (LEHR) and low energy all-purpose (LEAP) collimators were used for the Symbia TruePoint. For animal imaging, 16.8 MBq of 125 I was administered to BALB/c mice intravenously, and the planar image and single-photon emission computed tomography (SPECT) were obtained using both scanners. Results: The resolution of 125I for the Inveon scanner was 3.98 mm full width at half maximum (FWHM) at a 30-mm distance with the MHR-0.5 collimator, and the value of Symbia scanner was 8.72 mm FWHM at a 30-mm distance with the LEHR collimator. The sensitivity of 125 I for the Inveon scanner was 21.87 cps/MBq, and the value for the clinical scanner was 30.55 cps/MBq. The planar images of mice were successfully obtained at the level of evaluating specific binding in both scanners. Conclusion: 125 I small animal imaging can be achieved with a clinical scanner. This result may enhance the utilization of 125 I small animal imaging using a clinical scanner.
- Beekman, F.J., et al. (2002) Towards in Vivo Nuclear Microscopy: Iodine-125 Imaging in Mice Using Micro-Pinholes. European Journal of Nuclear Medicine and Molecular Imaging, 29, 933-938. http://dx.doi.org/10.1007/s00259-002-0805-6
- Dilmanian, F.A., et al. (1990) A High Resolution SPECT System Based on a Microchannel-Plate Imagier. IEEE Transactions on Nuclear Science, 37, 687-695. http://dx.doi.org/10.1109/23.106699
- Kim, E.M., et al. (2009) Characterization, Biodistribution and Small-Animal SPECT of I-125-Labeled c-Met Binding Peptide in Mice Bearing c-Met Receptor Tyrosine Kinase-Positive Tumor Xenografts. Nuclear Medicine and Biology, 36, 371-378. http://dx.doi.org/10.1016/j.nucmedbio.2009.01.005
- Franc, B.L., et al. (2008) Small-Animal SPECT and SPECT/CT: Important Tools for Preclinical Investigation. Journal of Nuclear Medicine, 49, 1651-1663. http://dx.doi.org/10.2967/jnumed.108.055442
- Meikle, S.R., et al. (2005) Small Animal SPECT and Its Place in the Matrix of Molecular Imaging Technologies. Physics in Medicine and Biology, 50, R45-R61. http://dx.doi.org/10.1088/0031-9155/50/22/r01
- Weber, D.A., et al. (1994) Pinhole SPECT: An Approach to in Vivo High Resolution SPECT Imaging in Small Laboratory Animals. Journal of Nuclear Medicine, 35, 342-348.
- Herschman, H.R. (2003) Molecular Imaging: Looking at Problems, Seeing Solutions. Science, 302, 605-608. http://dx.doi.org/10.1126/science.1090585
- Hwang, A.B., et al. (2008) Assessment of the Sources of Error Affecting the Quantitative Accuracy of SPECT Imaging in Small Animals. Physics in Medicine and Biology, 53, 2233-2252. http://dx.doi.org/10.1088/0031-9155/53/9/002
- Scherfler, C. and Decristoforo, C. (2006) Small Animal Imaging Using a Conventional Gamma Camera Exemplified in Studies on the Striatal Dopaminergic System. Nuclear Medicine Review Central East Europe, 9, 6-11.
- Aide, N., et al. (2010) High Throughput Static and Dynamic Small Animal Imaging Using Clinical PET/CT: Potential Preclinical Applications. European Journal of Nuclear Medicine and Molecular Imaging, 37, 991-1001. http://dx.doi.org/10.1007/s00259-009-1352-1
- Khalil, M.M., Tremoleda, J.L., Bayomy, T.B. and Gsell, W. (2011) Molecular SPECT Imaging: An Overview. International Journal of Molecular Imaging, 2011, Article ID: 796025. http://dx.doi.org/10.1155/2011/796025
- Umeda, I.O., Tani, K., Tsuda, K., Kobayashi, M., Ogata, M., Kimura, S., et al. (2012) High Resolution SPECT Imaging for Visualization of Intratumoral Heterogeneity Using a SPECT/CT Scanner Dedicated for Small Animal Imaging. Annals of Nuclear Medicine, 26, 67-76. http://dx.doi.org/10.1007/s12149-011-0542-7