Evaluation of SON’OR<sup>©</sup>, a Medical Device for Provoked Otoacoustic Emissions and Brainstem Evoked Response Audiometry Made in Cameroon
- 1 Otorhinolaryngology Service, Centre Hospitalier d’Essos, Yaoundé, Cameroon
- 2 Regional Hospital Bertoua, Bertoua, Cameroon
- 3 BENDO Company, Yaoundé, Cameroon
- 4 General Hospital, Yaoundé, Cameroon
- 5 General Hospital, Douala, Cameroon
- 6 National Advanced School of Engineering, Yaoundé, Cameroon
- 7 Laquintinie Hospital, Douala, Cameroon
- 8 Cameroon Society of Otolaryngology and Head and Neck Surgery, Yaoundé, Cameroon
- 9 Chantal Biya International Research Center, Yaoundé, Cameroon
Abstract
Background: There is a huge gap in the audiological care in Africa by comparison with Western countries. Its main reason is the prohibitive cost of the medical devices used to screen or diagnose patients. A Cameroonian team tackled this problem by developing a medical device (SON’OR<sup> © </sup>) which integrates a new instrumentation amplifier structure dedicated to perform otoacoustic emissions (OAE) and brainstem evoked response audiometry (BERA). The major technical challenges to perform OAE and BERA are the synchronization and the amplification of signals of very low amplitude. In this work, we emphasize on the general criteria necessary and indispensable to achieve an optimal amplification. The application of a novel instrumentation amplifier structure characterized by its optimized noise factor in the case of BERA and OEA emissions provides simulations and experimental results fully in line with forecasts. The design of SON’OR <sup>©</sup> is based on general techniques of electronic instrumentation to which we associated the new instrumentation amplifier structure. Objective: To report the clinical evaluation of SON’OR <sup>©</sup> as a screening and diagnostic tool. Methods: We conducted a cross sectional comparative study in Centre Hospitalier d’Essos in Yaoundé. We tested SON’OR <sup>©</sup> on two sets of subjects, one for OAE with OTODYNAMICS Echoport ILO 292-II as gold standard and the other for BERA with NEUROSOFT NEURO AUDIO <sup>©</sup> as gold standard. Each patient was tested with both devices and then we studied the inter device differences and calculated the sensitivity, specificity, positive predictive value and negative predictive value for each test. Results: We got 52 subjects for OEA and 51 for BERA testing. Sex ratio was 1 woman for 2 men in both groups. Mean age was 24.86(SD = 10.53) and 26.33 (10.55) for OAE and BERA groups respectively. SON’OR <sup>©</sup> had good performances, showing sensitivity of 92.85%/95%, specificity of 96.77%/90.47%, positive predictive value of 95.21%/93.44% and negative predictive value of 95.23%/92.68% for OAE/ BERA respectively. Conclusion: SON’OR <sup>©</sup> has good characteristics as a medical diagnostic tool. Furthermore its stability and performances in poor electrical conditions make it a robust device really suited for resource limited settings.
- World Health Organization (2017) Deafness and Hearing Loss. http://www.who.int/mediacentre/factsheets/fs300/en/
- Northern, J.L. and Hayes, D. (1994) Universal Screening for Infant Hearing Impairment: Necessary, Beneficial and Justifiable. Audiology Today, 6, 1-4.
- Nagapoornima, P., Ramesh, A., Rao, S., Patricia, P.L., Gore, M., Dominic, M., et al. (2007) Universal Hearing Screening. Indian Journal of Pediatrics, 74, 545-549. https://doi.org/10.1007/s12098-007-0105-z
- Govaerts, P.J., De Beukelaer, C., Daemers, K., De Ceulaer, G., Yperman, M., Somers, T., et al. (2002) Outcome of Cochlear Implantation at Different Ages from 0 to 6 Years. Otology & Neurotology, 23, 885-890. https://doi.org/10.1097/00129492-200211000-00013
- Downs, M.P. and Yoshinaga-Itano, C. (1999) The Efficacy of Early Identification and Intervention for Children with Hearing Impairment. Pediatric Clinics of North America, 46, 79-87.
- Yoshinaga-Itano, C. (2003) Early Intervention after Universal Neonatal Hearing Screening: Impact on Outcomes. Mental Retardation and Developmental Disabilities Research Reviews, 9, 252-266. https://doi.org/10.1002/mrdd.10088
- Yoshinaga-Itano, C., Sedey, A.L., Coulter, D.K. and Mehl, A.L. (1998) Language of Early- and Later-Identified Children with Hearing Loss. Pediatrics, 102, 1161-1171. https://doi.org/10.1542/peds.102.5.1161
- Swanepoel, D.C.D., Delport, S.D. and Swart, J.G. (2004) Universal Newborn Hearing Screening in South Africa—A First-World Dream? South African Medical Journal, 94, 634-635.
- Ngounou, G. and Kom, M. (2014) Optimization of Noise in Non-Integrated Instrumentation Amplifier for the Amplification of Very Low Electrophysiological Signals. Case of Electrocardiographic Signals (ECG). Journal of Medical Systems, 38, 152. https://doi.org/10.1007/s10916-014-0152-8
- Verhoeven, C.J.M., van Staveren, A., Monna, G.L.E., Kouwenhoven, M.H.L. and Yildiz, E. (2003) Noise. Springer, Berlin, 85-150.
- Shojaei-Baghini, M., Lal, R.K. and Sharma, D.K. (2004) An Ultra Low-Power CMOS Instrumentation Amplifier for Biomedical Applications. IEEE International Workshop on Biomedical Circuits and Systems, Singapore, 1-3 December 2004, S1. https://doi.org/10.1109/BIOCAS.2004.1454082
- Koli, K. and Halonen, K.A. (2000) CMRR Enhancement Techniques for Current-Mode Instrumentation Amplifiers. IEEE Transactions on Circuits and Systems: Fundamental Theory and Applications, 47, 622-632. https://doi.org/10.1109/81.847869