Electrocardiographic interference and conductance volume measurements
- 1
- 2
- 3
- 4
Abstract
The conductance catheter technique enables continuous ventricular volume measurements based on the electrical conductance of blood within the ventricular cavity. However, ventricular excitation also produces a measurable electrical signal within the ventricular cavity. This study was undertaken to investigate the relationship between the ventricular electrogram and conductance volume measurements in a physical model of the left ventricle without parallel conductance. The ventricular electrogram was simulated with an ECG signal, ECGinput connected to two ring electrodes within the model ventricle. Conductance volume measurements were made with and without ECGinput. The difference between these measurements, GECG(t), represented the conductance volume due to ECGinput. GECG(t) varied as a function of the first-derivative of ECGinput with respect to time (r2=0.92, P<0.001). GECG(t), This primarily affected volume measurements during ventricular depolarisation; during this phase the volume measurement error varied widely between –12% and +9%. As a re-sult, end-diastole could not be reliably identified on the pressure-volume loop. The accuracy of conduc-tance volume measurements during late diastole and early isovolumic contraction are substantially affected by the ventricular electrogram. This may result in a significant error in end-diastolic volume estimates, which has important implications for the quantitative assessment of ventricular function including, in particular, the assessment of chamber compliance.
- Baan, J., Aouw Jong, T. T., Kerkhof, P. L., Moene, R. J., van Dijk, A. D., van der Velde, E. T., and Koops, J. (1981) Con-tinuous stroke volume and cardiac output from intra- ventricu-lar dimensions obtained with an impedance ca- theter, Cardio-vascular Research, 15, 328–334.
- Mur, G. and Baan, J., (1984) Computation of the input imped-ances of a catheter for cardiac volumetry, IEEE Transactions on Biomedical Engineering, 31, 448–453.
- Kass, D. A., (1992) Clinical evaluation of left heart function by conductance catheter technique, European Heart Journal, 13(Suppl E), 57–64.
- Wu, C. C., Skalak, T. C., Schwenk, T. R., Mahler, C. M., Anne, A., Finnerty, P. W., Haber, H. L., Weikle II, R. M. and Feldman, M. D., (1997) Accuracy of the conductance catheter for meas-urement of ventricular volumes seen clinically: Effects of elec-tric field homogeneity and parallel conductance, IEEE Trans-actions on Biomedical Engineering, 44, 266–277.
- Salo, R. W., (1989) The theoretical basis of a compartmental model for the determination of volume by impedance, Automedica, 11, 299–310.
- Wei, C. L., Valvano, J. W., Feldman, M. D., and Pearce, J. A., (2005) Nonlinear conductance-volume relationship for murine conductance catheter measurement system, IEEE Transactions on Biomedical Engineering, 52, 654– 661.
- Salo, R. W., Wallner, T. G., and Pederson, B. D., (1986) Meas-urement of ventricular volume by intracardiac impedance: Theoretical and empirical approaches, IEEE Transactions on Biomedical Engineering, 33, 189?195.
- Baan, J., van der Velde, E. T., de Bruin, H. G., Smeenk, G. J., Koops, J., van Dijk, A. D., Temmerman, D., Senden, J., and Buis, B., (1984) Continuous measurement of left ventricular volume in animals and humans by conductance catheter, Cir-culation, 70, 812–823.
- Al-Khalidi, A. H., Townend, J. N., Bonser, R. S., and Coote, J. H., (1998) Validation of the conductance catheter method for measurement of ventricular volumes under varying conditions relevant to cardiac surgery, Am- erican Journal of Cardiology, 82, 1248–1252.
- McGuirk, S. P., Barron, D. J., Ewert, D., and Coote, J. H., (2009) Calibrating volume measurements made using the dual-field conductance catheter, Journal of Biomedical Science and Engineering, In Press.
- Durrer, D., Van Dam, R. T., Freud, G. E., Janse, M. J., Meijler, F. L., and Arzbaecher, R. C., (1970) Total excitation of the isolated human heart, Circulation, 41, 899– 912.