Bioreactivity of Stent Material: <i>In Vitro</i> Impact of New Twinning-Induced Plasticity Steel on Platelet Activation — Oak Academic Publishing
Research ArticleOpen AccessGoogle Scholar indexed
Bioreactivity of Stent Material: <i>In Vitro</i> Impact of New Twinning-Induced Plasticity Steel on Platelet Activation
Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Plateforme de Cytométrie de Flux, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Division of Cardiology, Cliniques Universitaires Saint-Luc, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Department of Materials and Process Engineering, Institute of Mechanics, Materials and Civil Engineering, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Division of Cardiology, Cliniques Universitaires Saint-Luc, Université catholique de Louvain (UCLouvain), Brussels, Belgium
,
Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
1 Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
2 Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
3 Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
4 Plateforme de Cytométrie de Flux, Université catholique de Louvain (UCLouvain), Brussels, Belgium
5 Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
6 Division of Cardiology, Cliniques Universitaires Saint-Luc, Université catholique de Louvain (UCLouvain), Brussels, Belgium
7 Department of Materials and Process Engineering, Institute of Mechanics, Materials and Civil Engineering, Université catholique de Louvain (UCLouvain), Brussels, Belgium
8 Division of Cardiology, Cliniques Universitaires Saint-Luc, Université catholique de Louvain (UCLouvain), Brussels, Belgium
9 Pôle de Recherche Cardiovasculaire, Université catholique de Louvain (UCLouvain), Brussels, Belgium
A current challenge concerns developing new bioresorbable stents that combine optimal mechanical properties and biodegradation rates with limited thrombogenicity. In this context, twinning-induced plasticity (TWIP) steels are good material candidates. In this work, the hemocompatibility of a new TWIP steel was studied in vitro via hemolysis and platelet activation assessments. Cobalt chromium (CoCr) L605 alloy, pure iron (Fe), and magnesium (Mg) WE43 alloy were similarly studied for comparison. No hemolysis was induced by TWIP steel, pure Fe, or L605 alloy. Moreover, L605 alloy did not affect CD62P exposure, α IIb β 3 activation at the platelet surface, or phosphorylation of protein kinase C (PKC) substrates upon thrombin stimulation. In contrast, TWIP steel and pure Fe significantly decreased platelet response to the agonist. Given that similar inhibitory effects were obtained when using a conditioned medium previously incubated with TWIP steel, we postulated TWIP steel corrosion to be likely to release components counteracting platelet activation. We showed that the main ion form present in the conditioned medium is Fe 3+ . In conclusion, TWIP steel resorbable scaffold displays anti-thrombogenic properties in vitro , which suggests that it could be a promising platform for next-generation stent technologies.
Libby, P., Ridker, P.M. and Hansson, G.K. (2011) Progress and Challenges in Translating the Biology of Atherosclerosis. Nature, 473, 317-325. https://doi.org/10.1038/nature10146
Corban, M.T., Lerman, L.O. and Lerman, A. (2019) Endothelial Dysfunction. Arteriosclerosis, Thrombosis, and Vascular Biology, 39, 1272-1274. https://doi.org/10.1161/ATVBAHA.119.312836
Maddox, T.M., Stanislawski, M.A., Grunwald, G.K., Bradley, S.M., Ho, P.M., Tsai, T.T., Patel, M.R., Sandhu, A., Valle, J., Magid, D.J., Leon, B., Bhatt, D.L., Fihn, S.D. and Rumsfeld, J.S. (2014) Nonobstructive Coronary Artery Disease and Risk of Myocardial Infarction. JAMA, 312, 1754-1763. https://doi.org/10.1001/jama.2014.14681
Du, F. and Zhou, J. (2018) Vascular Intervention: From Angioplasty to Bioresorbable Vascular Scaffold. Advances in Experimental Medicine and Biology, 1097, 181-189. https://doi.org/10.1007/978-3-319-96445-4_9
Iqbal, J., Gunn, J. and Serruys, P.W. (2013) Coronary Stents: Historical Development, Current Status and Future Directions. British Medical Bulletin, 106, 193-211. https://doi.org/10.1093/bmb/ldt009
Borhani, S., Hassanajili, S., Ahmadi Tafti, S.H. and Rabbani, S. (2018) Cardiovascular Stents: Overview, Evolution, and Next Generation. Progress in Biomaterials, 7, 175-205. https://doi.org/10.1007/s40204-018-0097-y
Weber, M., Steinle, H., Golombek, S., Hann, L., Schlensak, C., Wendel, H.P. and Avci-Adali, M. (2018) Blood-Contacting Biomaterials: In Vitro Evaluation of the Hemocompatibility. Frontiers in Bioengineering and Biotechnology, 6, 99. https://doi.org/10.3389/fbioe.2018.00099
Yen, J.H., Chen, S.F., Chern, M.K. and Lu, P.C. (2014) The Effect of Turbulent Viscous Shear Stress on Red Blood Cell Hemolysis. Journal of Artificial Organs, 17, 178-185. https://doi.org/10.1007/s10047-014-0755-3
Merle, N.S., Grunenwald, A., Figueres, M.L., Chauvet, S., Daugan, M., Knockaert, S., Robe-Rybkine, T., Noe, R., May, O., Frimat, M., Brinkman, N., Gentinetta, T., Miescher, S., Houillier, P., Legros, V., Gonnet, F., Blanc-Brude, O.P., Rabant, M., Daniel, R., Dimitrov, J.D. and Roumenina, L.T. (2018) Characterization of Renal Injury and Inflammation in an Experimental Model of Intravascular Hemolysis. Frontiers in Immunology, 9, 179. https://doi.org/10.3389/fimmu.2018.00179
Rossaint, J., Margraf, A. and Zarbock, A. (2018) Role of Platelets in Leukocyte Recruitment and Resolution of Inflammation. Frontiers in Immunology, 9, 2712. https://doi.org/10.3389/fimmu.2018.02712
Etulain, J. and Schattner, M. (2014) Glycobiology of Platelet-Endothelial Cell Interactions. Glycobiology, 24, 1252-1259. https://doi.org/10.1093/glycob/cwu056
Diamond, S.L. (2016) Systems Analysis of Thrombus Formation. Circulation Research, 118, 1348-1362. https://doi.org/10.1161/CIRCRESAHA.115.306824
Jaffer, I.H., Fredenburgh, J.C., Hirsh, J. and Weitz, J.I. (2015) Medical Device-Induced Thrombosis: What Causes It and How Can We Prevent It? Journal of Thrombosis and Haemostasis, 13, S72-S81. https://doi.org/10.1111/jth.12961
Swieringa, F., Spronk, H.M.H., Heemskerk, J.W.M. and van der Meijden, P.E.J. (2018) Integrating Platelet and Coagulation Activation in Fibrin Clot Formation. Research and Practice in Thrombosis and Haemostasis, 2, 450-460. https://doi.org/10.1002/rth2.12107
Dziewierz, A. and Dudek, D. (2018) Current Perspectives on the Role of Bioresorbable Scaffolds in the Management of Coronary Artery Disease. Kardiologia Polska, 76, 1043-1054. https://doi.org/10.5603/KP.a2018.0130
Dave, B. (2016) Bioresorbable Scaffolds: Current Evidences in the Treatment of Coronary Artery Disease. Journal of Clinical and Diagnostic Research, 10, OE01-OE07. https://doi.org/10.7860/JCDR/2016/21915.8429
Moravej, M. and Mantovani, D. (2011) Biodegradable Metals for Cardiovascular Stent Application: Interests and New Opportunities. International Journal of Molecular Sciences, 12, 4250-4270. https://doi.org/10.3390/ijms12074250
Walker, E.K., Nauman, E.A., Allain, J.P. and Stanciu, L.A. (2015) An in Vitro Model for Preclinical Testing of Thrombogenicity of Resorbable Metallic Stents. Journal of Biomedical Materials Research Part A, 103, 2118-2125. https://doi.org/10.1002/jbm.a.35348
Liu, B. and Zheng, Y.F. (2011) Effects of Alloying Elements (Mn, Co, Al, W, Sn, B, C and S) on Biodegradability and in Vitro Biocompatibility of Pure Iron. Acta Biomaterialia, 7, 1407-1420. https://doi.org/10.1016/j.actbio.2010.11.001
Liu, Y., Wu, Y., Bian, D., Gao, S., Leeflang, S., Guo, H., Zheng, Y. and Zhou, J. (2017) Study on the Mg-Li-Zn Ternary Alloy System with Improved Mechanical Properties, Good Degradation Performance and Different Responses to Cells. Acta Biomaterialia, 62, 418-433. https://doi.org/10.1016/j.actbio.2017.08.021
Yahata, C. and Mochizuki, A. (2017) Platelet Compatibility of Magnesium Alloys. Materials Science & Engineering C—Materials for Biological Applications, 78, 1119-1124. https://doi.org/10.1016/j.msec.2017.04.153
Feyerabend, F., Wendel, H.P., Mihailova, B., Heidrich, S., Agha, N.A., Bismayer, U. and Willumeit-Romer, R. (2015) Blood Compatibility of Magnesium and Its Alloys. Acta Biomaterialia, 25, 384-394. https://doi.org/10.1016/j.actbio.2015.07.029
Peuster, M., Hesse, C., Schloo, T., Fink, C., Beerbaum, P. and von Schnakenburg, C. (2006) Long-Term Biocompatibility of a Corrodible Peripheral Iron Stent in the Porcine Descending Aorta. Biomaterials, 27, 4955-4962. https://doi.org/10.1016/j.biomaterials.2006.05.029
Peuster, M., Wohlsein, P., Brugmann, M., Ehlerding, M., Seidler, K., Fink, C., Brauer, H., Fischer, A. and Hausdorf, G. (2001) A Novel Approach to Temporary Stenting: Degradable Cardiovascular Stents Produced from Corrodible Metal-Results 6-18 Months after Implantation into New Zealand White Rabbits. Heart, 86, 563-569. https://doi.org/10.1136/heart.86.5.563
Waksman, R., Pakala, R., Baffour, R., Seabron, R., Hellinga, D. and Tio, F.O. (2008) Short-Term Effects of Biocorrodible Iron Stents in Porcine Coronary Arteries. Journal of Interventional Cardiology, 21, 15-20. https://doi.org/10.1111/j.1540-8183.2007.00319.x
Haude, M., Erbel, R., Erne, P., Verheye, S., Degen, H., Vermeersch, P., Weissman, N., Prati, F., Bruining, N., Waksman, R. and Koolen, J. (2016) Safety and Performance of the Drug-Eluting Absorbable Metal Scaffold (DREAMS) in Patients with De Novo Coronary Lesions: 3-Year Results of the Prospective, Multicentre, First-in-Man BIOSOLVE-I Trial. EuroIntervention, 12, e160-e166. https://doi.org/10.4244/EIJ-D-15-00371
Sakamoto, A., Jinnouchi, H., Torii, S., Virmani, R. and Finn, A.V. (2018) Understanding the Impact of Stent and Scaffold Material and Strut Design on Coronary Artery Thrombosis from the Basic and Clinical Points of View. Bioengineering (Basel), 5, 71. https://doi.org/10.3390/bioengineering5030071
Sezer, N., Evis, Z., Kayhan, S.M., Tahmasebifar, A. and Koç, M. (2018) Review of Magnesium-Based Biomaterials and Their Applications. Journal of Magnesium and Alloys, 6, 23-43. https://doi.org/10.1016/j.jma.2018.02.003
Idrissi, H., Renard, K., Ryelandt, L., Schryvers, D. and Jacques, P.J. (2010) On the Mechanism of Twin Formation in Fe-Mn-C TWIP Steels. Acta Materialia, 58, 2464-2476. https://doi.org/10.1016/j.actamat.2009.12.032
Renard, K. and Jacques, P.J. (2012) On the Relationship between Work Hardening and Twinning Rate in TWIP Steels. Materials Science and Engineering A, 542, 8-14. https://doi.org/10.1016/j.msea.2012.01.123
De Cooman, B.C., Estrin, Y. and Kim, S.K. (2018) Twinning-Induced Plasticity (TWIP) Steels. Acta Materialia, 142, 283-362. https://doi.org/10.1016/j.actamat.2017.06.046
Goodhead, L.K. and MacMillan, F.M. (2017) Measuring Osmosis and Hemolysis of Red Blood Cells. Advances in Physiology Education, 41, 298-305. https://doi.org/10.1152/advan.00083.2016
ASTM F. F756-08 (2013) Standard Practice for Assessment of Hemolytic Properties of Materials, ASTM Book of Standards. ASTM International, West Conshohocken.
Cheng, J., Liu, B., Wu, Y.H. and Zheng, Y.F. (2013) Comparative in Vitro Study on Pure Metals (Fe, Mn, Mg, Zn and W) as Biodegradable Metals. Journal of Materials Science & Technology, 29, 619-627. https://doi.org/10.1016/j.jmst.2013.03.019
Huang, T., Cheng, Y. and Zheng, Y. (2016) In Vitro Studies on Silver Implanted Pure Iron by Metal Vapor Vacuum Arc Technique. Colloids and Surfaces B: Biointerfaces, 142, 20-29. https://doi.org/10.1016/j.colsurfb.2016.01.065
Li, M., Cheng, Y., Zheng, Y.F., Zhang, X., Xi, T.F. and Wei, S.C. (2012) Surface Characteristics and Corrosion Behaviour of WE43 Magnesium Alloy Coated by SiC Film. Applied Surface Science, 258, 3074-3081. https://doi.org/10.1016/j.apsusc.2011.11.040
Wei, Z., Tian, P., Liu, X. and Zhou, B. (2014) Hemocompatibility and Selective Cell Fate of Polydopamine-Assisted Heparinized PEO/PLLA Composite Coating on Biodegradable AZ31 Alloy. Colloids and Surfaces B: Biointerfaces, 121, 451-460. https://doi.org/10.1016/j.colsurfb.2014.06.036
Zhou, W.R., Zheng, Y.F., Leeflang, M.A. and Zhou, J. (2013) Mechanical Property, Biocorrosion and in Vitro Biocompatibility Evaluations of Mg-Li-(Al)-(RE) Alloys for Future Cardiovascular Stent Application. Acta Biomaterialia, 9, 8488-8498. https://doi.org/10.1016/j.actbio.2013.01.032
Mao, L., Yuan, G., Niu, J., Zong, Y. and Ding, W. (2013) In Vitro Degradation Behavior and Biocompatibility of Mg-Nd-Zn-Zr Alloy by Hydrofluoric Acid Treatment. Materials Science & Engineering C-Materials for Biological Applications, 33, 242-250. https://doi.org/10.1016/j.msec.2012.08.036
Gu, X., Zheng, Y., Cheng, Y., Zhong, S. and Xi, T. (2009) In Vitro Corrosion and Biocompatibility of Binary Magnesium Alloys. Biomaterials, 30, 484-498. https://doi.org/10.1016/j.biomaterials.2008.10.021
Huang, T., Cheng, J., Bian, D. and Zheng, Y. (2016) Fe-Au and Fe-Ag Composites as Candidates for Biodegradable Stent Materials. Journal of Biomedical Materials Research Part B: Applied Biomaterials, 104, 225-240. https://doi.org/10.1002/jbm.b.33389
ISO 10993 (2002) Biological Evaluation of Medical Devices—Part 4: Selection of Tests for Interactions with Blood. International Organization for Standardization, Genève.
Moore, S.F., Hunter, R.W. and Hers, I. (2014) Protein Kinase C and P2Y12 Take Center Stage in Thrombin-Mediated Activation of Mammalian Target of Rapamycin Complex 1 in Human Platelets. Journal of Thrombosis and Haemostasis, 12, 748-760. https://doi.org/10.1111/jth.12552
Huang, T., Cheng, J. and Zheng, Y.F. (2014) In Vitro Degradation and Biocompatibility of Fe-Pd and Fe-Pt Composites Fabricated by Spark Plasma Sintering. Materials Science & Engineering C-Materials for Biological Applications, 35, 43-53. https://doi.org/10.1016/j.msec.2013.10.023
Hermawan, H., Purnama, A., Dube, D., Couet, J. and Mantovani, D. (2010) Fe-Mn Alloys for Metallic Biodegradable Stents: Degradation and Cell Viability Studies. Acta Biomaterialia, 6, 1852-1860. https://doi.org/10.1016/j.actbio.2009.11.025
Nemmar, A., Beegam, S., Yuvaraju, P., Yasin, J., Tariq, S., Attoub, S. and Ali, B.H. (2016) Ultrasmall Superparamagnetic Iron Oxide Nanoparticles Acutely Promote Thrombosis and Cardiac Oxidative Stress and DNA Damage in Mice. Particle and Fibre Toxicology, 13, 22. https://doi.org/10.1186/s12989-016-0132-x
Rooyakkers, T.M., Stroes, E.S., Kooistra, M.P., van Faassen, E.E., Hider, R.C., Rabelink, T.J. and Marx, J.J. (2002) Ferric Saccharate Induces Oxygen Radical Stress and Endothelial Dysfunction in Vivo. European Journal of Clinical Investigation, 32, 9-16. https://doi.org/10.1046/j.1365-2362.2002.0320s1009.x
Eckly, A., Hechler, B., Freund, M., Zerr, M., Cazenave, J.P., Lanza, F., Mangin, P.H. and Gachet, C. (2011) Mechanisms Underlying FeCl3-Induced Arterial Thrombosis. Journal of Thrombosis and Haemostasis, 9, 779-789. https://doi.org/10.1111/j.1538-7836.2011.04218.x
Ciciliano, J.C., Sakurai, Y., Myers, D.R., Fay, M.E., Hechler, B., Meeks, S., Li, R., Dixon, J.B., Lyon, L.A., Gachet, C. and Lam, W.A. (2015) Resolving the Multifaceted Mechanisms of the Ferric Chloride Thrombosis Model Using an Interdisciplinary Microfluidic Approach. Blood, 126, 817-824. https://doi.org/10.1182/blood-2015-02-628594
Schoenwaelder, S.M. and Jackson, S.P. (2015) Ferric Chloride Thrombosis Model: Unraveling the Vascular Effects of a Highly Corrosive Oxidant. Blood, 126, 2652-2653. https://doi.org/10.1182/blood-2015-09-668384
Barr, J.D., Chauhan, A.K., Schaeffer, G.V., Hansen, J.K. and Motto, D.G. (2013) Red Blood Cells Mediate the Onset of Thrombosis in the Ferric Chloride Murine Model. Blood, 121, 3733-3741. https://doi.org/10.1182/blood-2012-11-468983
Miron, V.R., Bauermann, L., Morsch, A.L., Zanin, R.F., Correa, M., da Silva, A.C., Mazzanti, C., Morsch, V.M., Lunkes, G.I. and Schetinger, M.R. (2007) Enhanced NTPDase and 5’-Nucleotidase Activities in Diabetes Mellitus and Iron-Overload Model. Molecular and Cellular Biochemistry, 298, 101-107. https://doi.org/10.1007/s11010-006-9357-6