Acuros XB Algorithm vs. Anisotropic Analytical Algorithm: A Dosimetric Study Using Heterogeneous Phantom and Computed Tomography (CT) Data Sets of Esophageal Cancer Patients
- 1 Department of Radiation Oncology, Arizona Center for Cancer Care, Peoria, USA
- 2 Department of Radiation Oncology, Arizona Center for Cancer Care, Peoria, USA
- 3 Premiere Oncology, Fort Myers, USA
- 4 Department of Radiation Oncology, Arizona Center for Cancer Care, Peoria, USA
- 5 Department of Radiation Oncology, Arizona Center for Cancer Care, Peoria, USA
- 6 Department of Radiation Oncology, Arizona Center for Cancer Care, Peoria, USA
Abstract
Our purpose in this study was to assess the dosimetric impact of the Acuros XB algorithm (AXB), in comparison with Anisotropic Analytical Algorithm (AAA) calculations, for esophageal cancer treatment plans created with RapidArc technique. First, we performed a phantom study by comparing the percent depth dose (PDD) calculated by AXB and AAA against the measured PDD in a slab phantom containing a 2 cmair gap thickness. Second, we performed a clinical study using a computed tomography (CT) data set from 10 esophageal cancer patients. The treatment plans calculated by AXB and AAA were evaluated for planning target volume (PTV) coverage, doses to the PTV and organs at risk (OARs). Dose calculations by the AXB and AAA were done for identical beam parameters. The AXB showed better agreement (within ±0.5%) with measurements than did the AAA ( ? 4.9% to ?6.2%). In comparison to the AAA, the AXB predicted a higher maximum PTV dose (2.0%), but lower mean (1.1%) and minimum (2.5%) PTV doses as well as reduced PTV coverage (9.1%). The averaged mean doses to all OARs predicted by the AXB were lower (up to 3.6%), and the percentage of lungs volume receiving at least 20 and 5 Gy were lower by about 3.6% in the AXB plans compared to the AAA plans. The AXB is more accurate than the AAA for dose predictions when air medium is involved. The use of AXB is more likely to avoid dose overestimation or underestimation for the esophageal cancer treatment plans compared to AAA.
- B. Minsky, T. Pajak, R. Ginsberg, T. Pisansky, J. Martenson, R. Komaki, G. Okawara, S. Rosenthal and D. Kelsen “INT 0123 (Radiation Therapy Oncology Group 94-05) Phase III Trial of Combined-Modality Therapy for Esophageal Cancer: High-Dose Versus Standard-Dose Radiation Therapy,” Journal of Clinical Oncology, Vol. 20, No. 5, 2002, pp. 1167-1174. doi:10.1200/JCO.20.5.1167
- C. C. Ling, P. Zhang, Y. Archambault, J. Bocanek, G. Tang and T. Losasso, “Commissioning and Quality Assurance of RapidArc Radiotherapy Delivery System,” Journal of Clinical Oncology, Vol. 72, No. 2, 2008, pp. 575-581. doi:10.1016/j.ijrobp.2008.05.060
- International Commission on Radiation Units and Measurements (ICRU), “Determination of Absorbed Dose in a Patient Irradiated by Beams of X and Gamma Rays in Radiotherapy Procedures,” ICRU Report, Bethesda, 1976.
- B. Fraas, J. Smathers and J. Deye, “Summary and Recommendations of a National Cancer Institute Workshop on Issues Limiting the Clinical Use of Monte Carlo Dose Calculation Algorithms for Megavoltage External Beam Radiation Therapy,” Medical Physics, Vol. 30, No. 12, 2003, pp. 3206-3216. doi:10.1118/1.1626990
- O. Vassiliev, T. Wareing, J. McGhee, G. Failla, M. Salehpour and F. Mourtada, “Validation of a New Grid Based Blotzmann Equation Solver for Dose Calculation in Radiotherapy with Photon Beams,” Physics in Medicine and Biology, Vol. 55, No. 3, 2010, pp. 581-598. doi:10.1088/0031-9155/55/3/002
- K. Bush, I. M. Gagne, S. Zavgorodni, W. Ansbacher and W. Beckham, “Dosimetric Validation of Acuros XB with Monte Carlo Methods for Photon Dose Calculations,” Medical Physics, Vol. 38, No. 4, 2011, pp. 2208-2221. doi:10.1118/1.3567146
- A. Fogliata, G. Nicolini, A. Clivio, E. Vanetti, P. Mancosu and L. Cozzi, “Dosimetric Validation of Acuros XB Advanced Dose Calculation Algorithm: Fundamental Characterization in Water,” Physics in Medicine and Biology, Vol. 56, No. 6, 2011, pp. 1879-1904. doi:10.1088/0031-9155/56/6/022
- L. Tillikainen, H. Helminen, T. Torsti, S. Siljamaki, J. Alakuijala, J. Pyyry and W. Ulmer, “3D Pencil-Beam-Based Superposition Algorithm for Photon Dose Calculation in Heterogeneous Media,” Physics in Medicine and Biology, Vol. 53, No. 14, 2008, pp. 3821-3839. doi:10.1088/0031-9155/53/14/008
- L. Tillikainen, S. Siljam?ki, H. Helminen, J. Alakuijala and J. Pyyry, “Determination of Parameters for a Multiple-Source Model of Megavoltage Photon Beams Using Optimization Methods,” Physics in Medicine and Biology, Vol. 52, No. 5, 2007, pp. 1441-1467. doi:10.1088/0031-9155/52/5/015