Impact of a Radio Frequency Electronic Article Surveillance (EAS) System on Active Implants
- 1 Institute of Health Care Engineering with European Notified Body #0636 of Medical Devices, Graz University of Technology, Graz, Austria
- 2 Institute of Health Care Engineering with European Notified Body #0636 of Medical Devices, Graz University of Technology, Graz, Austria
- 3 Institute of Health Care Engineering with European Notified Body #0636 of Medical Devices, Graz University of Technology, Graz, Austria
Abstract
In view of the omnipresence of electronic article surveillance (EAS) systems in daily life and the increasing number of patients with active implants, there is concern about adverse electromagnetic interference in particular cardiac pacemakers (CPM) and cardioverter defibrillators (ICD), which due to sensing electrocardial signals are particularly vulnerable. To provide quantitative information interference of monopolar CPM and ICD by EAS systems operating at 8.2MHz radiofrequency electromagnetic fields (EMF) investigations have been performed by exposing numerical anatomical models of pacemaker patients with implants at the conventional left or right pectoral sites and at the abdomen to magnetic fields of a simulated EAS gate source. Investigation of normal position in the centre and worst case with the back next to the gate showed that adverse interference such as inadequate sensing need not be expected at any position. This applies for conventional sensing thresholds even if the exposure span of existing EAS systems is taken into account. However, if full use is made of the newly expanded exposure budget, adverse interference cannot be excluded.
- J. H. Bernhardt, A. F. McKinlay, R. Matthes, et al., “Possible Health Risk to the General Public from the Use of Security and Similar Devices,” International Commission on Non-Ionizing Radiation Protection, Munic, 2002.
- W. Boivin, J. Coletta and L. Kerr, “Characterization of the Magnetic Fields around Walk-Through and HandHeld Metal Detectors,” Health Physics, Vol. 84, No. 5, 2003, pp. 582-593. doi:10.1097/00004032-200305000-00003
- C. Harris, W. Boivin, S. Boyd, J. Coletta, L. Kerr, K. Kempa and S. Aronoiw, “Electromagnetic Field Strength Levels Surrounding Electronic Article Surveillance (EAS) Systems,” Health Physics, Vol. 78, No. 1, 2000, pp. 2127. doi:10.1097/00004032-200001000-00005
- M. Martinéz-Búrdalo, A. Sanchis, A. Matín and R. Villar, “Comparison of SAR and Induced Current Densities in Adults and Children Exposed to Electromagnetic Fields from Electronic Article Surveillance Devices,” Physics in Medicine and Biology, Vol. 55, No. 4, 2010, pp. 10411055. doi:10.1088/0031-9155/55/4/009
- G. Neubauer, H. Molla-Djaffari, K. D. Pühringer, H. Garn, N. Winkler, H. Prei? and G. Schmid, “Measurement and Safety Assessment of Electromagentic Fields around Anti-Theft Devices (German),” AUVA Report #23,Vienna, 1998.
- J. Trulsson, G. Anger and U. Estenberg, “Assessment of Magnetic Fields Surrounding Electronic Article Surveillance Systems in Sweden,” Bioelectromagnetics, Vol. 28, No. 8, 2007, pp. 664-666. doi:10.1002/bem.20359
- S. J. Seidman, R. Brockman, B. M. Lewis, J. Guag, M. J. Shein, W. J. Clement, J. Kippola, D. Digby, C. Barber and D. Huntwork, “In Vitro Tests Reveqal Sample Radiofrequency Identification Readers Inducing Clinically Significant Electgromagnetic Interference to Implantable Pacemakers and Implantable Cardioverter-Defibrillators,” Heart Rhythm, Vol. 7, No. 1, 2010, pp. 99-107. doi:10.1016/j.hrthm.2009.09.071
- W. Irnich, “Electronic Security Systems and active Implantable Medical Devices,” Pacing and Clinical Electrophysiology, Vol. 25, No. 8, 2002, pp. 1235-1258. doi:10.1046/j.1460-9592.2002.01235.x
- R. Frank, “Behaviour of 20 Pacemakers as Patients Pass through 2 Models of Theft-Detection Doors (French),” Ann Cardiol Angiol, Vol. 49, No. 3, 2000, pp. 187-197.
- M. E. McIvor, J. Reddinger, E. Floden and R. C. Sheppard, “Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS),” Pacing and Clinical Electrophysiology, Vol. 21, No. 10, 1998, pp. 1847-1861. doi:10.1111/j.1540-8159.1998.tb00002.x