Scanning Electron Micrographic Investigation of Explanted (Used) Metallic Biomedical Implants for Corrosion and Other Mechanical Damage Responsible for Failure
- 1 Department of Mechanical Engineering, IFTM University, Moradabad, India
- 2 Department of Mechanical Engineering, IFTM University, Moradabad, India
- 3 Middle East College, Knowledge Oasis Muscat, Muscat, Sultanate of Oman
Abstract
This paper deals with detailed corrosion analysis of explanted devices. The study of total 6 different types of orthopedic metallic implant was carried out after collecting the clinical report from the doctors, who performed these implantations. The clinical report covered the purpose of implantation, body part where implantation was done, and physiological reasons of removal of implant. The metallurgical investigation to study corrosion and any other mechanical damage to the implant surface during their service period was done using the Scanning Electron Micrography. SEM presented in this paper reveals the presence of in- vi tr o corrosion and mechanical damage as well, which are corroborating well with clinical reports.
- Fraker, A.C. (1992) Corrosion, ASM Handbook. ASM International, 13, 1324-1335.
- Speirs, A.D., Nganbe, M., Louati, H, Khan, U. and Beaulé, P.E. (2010) In Vitro and Retrieval Analysis of Modular Necks in Total Hip Arthroplasty. Proceedings of 56th Annual Meeting of the Orthopaedic Research Society, McKay Orthopaedic Laboratory, University of Pennsylvania, Philadelphia, 6-9 March 2010, pp.
- Gonzalez, J.E.G. and Mirza-Rosca, J.C. (1999) Study of the Corrosion Behavior of Titanium and Some of Its Alloys for Biomedical and Dental Implant Applications. Journal of Electroanalytical Chemistry, 471, 109-115. http://dx.doi.org/10.1016/S0022-0728(99)00260-0
- Kannan, S., Balamurugan, A. and Rajeswari, S. (2005) Electrochemical Characterization of Hydroxyapatite Coatings on HNO3 Passivated 316L SS for Implant Applications. Electrochimicaacta, 50, 2065-2072. http://dx.doi.org/10.1016/j.electacta.2004.09.015
- Fossati, A., Borgioli, F., Galvanetto, E. and Bacci, T. (2006) Glow Discharge Nitriding of AISI 316L Austenitic Stainless Steel: Influence of Treatment Time. Surface & Coatings Technology, 200, 3511-3517. http://dx.doi.org/10.1016/j.surfcoat.2004.10.122
- Syrett, B.C. and Wing, S.S. (1978) An Electrochemical Investigation of Fretting Corrosion of Surgical Implant Materials. Corrosion, 34, 378-386. http://dx.doi.org/10.5006/0010-9312-34.11.378
- Rodrigues, D.C., Valderrama, P., Wilson Jr., T.G., Palmer, K., Thomas, A., Sridhar, S., Adapalli, A., Burbano, M. and Wadhwani, C. (2013) Titanium Corrosion Mechanisms in the Oral Environment: A Retrieval Study. Materials, 6, 5258-5274. http://dx.doi.org/10.3390/ma6115258
- Sargeant, A. and Goswami, T. (2007) Hip Impants—Ion concentrations. Materials & Design, 28, 155-171. http://dx.doi.org/10.1016/j.matdes.2005.05.018
- Cheng, Y., Hu, J., Zhang, C., Wang, Z., Hao, Y. and Gao, B. (2013) Corrosion Behavior of Novel Ti-24Nb-4Zr-7.9Sn Alloy for Dental Implant Applications in Vitro. Journal of Biomedical Materials Research Part B, 101, 287-294. http://dx.doi.org/10.1002/jbm.b.32838
- Waddell, J.N., Payne, A.G.T. and Kieser, J.A. (2010) Scanning Electron Microscopy Observations of Failures of Implant Overdenture Bars: A Case Series Report. Clinical Implant Dentistry and Related Research, 12, 26-38. http://dx.doi.org/10.1111/j.1708-8208.2008.00127.x
- Rodrigues, D.C., Urban, R.M., Jacobs, J.J. and Gilbert, J.L. (2009) In Vivo Severe Corrosion and Hydrogen Embrittlement of Retrieved Modular Body Titanium Alloy Hip-Implants. Journal of Biomedical Materials Research Part B: Applied Biomaterials, 88, 206-219. http://dx.doi.org/10.1002/jbm.b.31171