Clinical Analysis of a Coaxial Dilator Set Attached to Needle Puncture for Percutaneous Nephrolithotomy
- 1 Master’s Program of Post Graduation in Surgery, Center of Health Sciences, Federal University of Pernambuco, Pernambuco, Brazil
- 2 Department of Surgery, Federal University of Pernambuco, Pernambuco, Brazil
- 3 Urology Service of the Hospital das Clinicas, Federal University of Pernambuco, Pernambuco, Brazil
- 4 Member of the Brazilian Society of Urology, Recife, Brazil
- 5 Member of the Brazilian Society of Urology, Recife, Brazil
Abstract
Introduction : Percutaneous approach to the kidney is a very useful alternative in renal surgery which minimizes the morbidity of a variety of procedures. Objectives : To evaluate the efficacy, practicity of a reusable system of metallic coaxial dilator coupled to a special puncture needle to perform lumbar puncture and dilation of the percutaneous tract. Methods : A randomized experimental study was carried out involving 50 individuals . These patients were randomized as follows: Group 1 had the procedure performed with the conventional disposable Amplatz set traditionally used for this procedure. Patients in Group 2 were operated utilizing the new reusable coaxial set specially designed for this purpose. The following parameters were measured to compare the 2 groups: sex Corporeal Mass Index (CMI), Pre and postoperative hematocrit and hemoglobin. Time interval between the puncture, dilation and access to the renal pelvis and insertion of the nephroscope were also analyzed. Results : There was no statistical difference between the 2 groups concerning hematocrit and hemoglobin changes when comparing pre with postoperative period. Patients in Group 2 required a significant lower time between puncture and final access to the upper collecting system. Conclusion : The new coaxial dilator set show ed to be as safe as the conventional Amplatz set with the advantage of reusability and decreas e of tract dilation time.
- H. Wilkinson, “Clinical Investigation and Management of Patients with Renal Stones,” Annals of Clinical Biochemistry, Vol. 38, No. 3, 2001, pp. 180-187. http://dx.doi.org/10.1258/000456 3011900623
- Ministério da Saúde, “Sistema de Informacoes Hospitalares do SUS (SIH/SUS),” Morbidade Hospital do SUS por Local de Internacao, Brasil, 2012. http://tabnet.datasus.gov.br/cgi/tabcgi.exe?sih/cnv/niuf.def
- T. Akman, M. Binbay, E. Tekinarslan, U. Ozkuvanci, C. Kezer, A. Erbin, Y. Berberoglu and A. Yaser-Muslumanoglu, “Outcomes of Percutaneous Nephrolithotomy in Patients with Solitary Kidney: A Single-Center Experience,” Urology, Vol. 78, No. 2, 2011, pp. 272-276. http://dx.doi.org/10.1016/ j.urology.2010.12.029
- M. S. Michel, L. Trojan and J. J. Rassweiles, “Complications in Percutaneous Nephrolithotomy,” European Urology, Vol. 51, No. 4, 2007, pp. 899-906. http://dx.doi.org/10.1016/j.eururo.2006.10.020
- T. Lopes, K. Sangam, P. Alken, B. S. Barroilhet, C. Saussine, L. Shi and J. de la Rossete, “Clinical Reserch Office of the Endourological Society. Percutaneous Nephrolithotomy Study Group,” Journal of Endourology, Vol. 25, No. 5, 2011, pp. 55-62. http://dx.doi.org/10.1089/end.2010.0488
- T. S. Shin, H. J. Cho, S. H. Hong, J. Y. Lee, S. W. Kim and T. K. Hwang, “Complications of Percutaneous Nephrolithotomy Classified the Modified Clavien Grading System: A Single Centes’s Experience over 16 years,” Korean Journal of Urology, Vol. 52, No. 11, 2011, pp. 769-775. http://dx.doi.org/ 10.4111/kju.2011.52.11.769
- Z. Huang, F. Fu, Z. Zhong, L. Zhang, R. Xu and X. Zhao, “Chinese Minimally Invasive Percutaneous Nephrolithotomy for Intrarenal Stones in Patients with Solitary Kidney: A Single-Center Experience,” PLoS One, Vol. 7, No. 7, 2012, Article ID: e40577. http://dx.doi.org/10.1371/journal.pone.0040577
- Y. B. Jeog, S. B. Jeon, Y. G. Kim and Y. K. Park, “POS01.97: Comparison of Balloon and Amplatz Fascial Dilation Systems for Percutaneous Tract Dilation in Percutaneous Nephrolithotomy,” Urology, Vol. 70, No. 3, 2007, pp. 219-220.
- M. L. Stoller, J. S. Wolf Jr. and M. A. Lezin, “Estimated Blood Loss and Transfusion Rate Associated with Percutaneous Nephrolithotomy,” Journal of Urology, Vol. 132, No. 1, 1994, pp. 1977-1981.
- A. Yamaguchi, A. Skolaricos, N. P. Buchholz, G. B. Chomón, M. Gasso, P. Saba, S. Nakada and J. de la Rossete, “Operating Times and Bleeding Complications in Percutaneous Nephrolithotomy: A Comparissonnof Tract Dilation Methods in 5.537 Patients in the Clinical Research Office of the Endourological Society Percutaneous Nephrolithotomy Global Study,” Journal—Endourological Society, Vol. 25, No. 6, 2011, pp. 933-9. http://dx.doi.org/10.1089/end.2010.0606